ELECTRICAL |
Input Power :P905 and 950S Combination |
+24 VDC, 950mA and -24VDC, 950mA |
| 950S Heater option |
115 VAC, 150 Watts |
| |
|
Infrared |
|
Range: |
Standard, 180° to 1650° C, 356° to 3000° F.
Optional extended range to 2500° C, 4530° F. |
| |
Accuracy |
Calibrated to +/-1% over full temperature range. |
| |
Resolution: |
.05 Degree internal. 1 Degree on P905 Remote Panel. |
|
TEMPERATURE |
Repeatability: |
< 1% over full range. |
|
MEASUREMENT |
|
Range: |
-50° to 1250° C, -58° to 2282° F. |
| |
Linearization: |
Thermocouple table loaded in non-volatile memory. |
|
Thermocouple |
Accuracy: |
Calibrated to +/-1% over full temperature range
using traceable K T/C. |
| |
Resolution: |
.02 Degree internal. 1 Degree on P905 Remote Panel. |
| |
Repeatability: |
< 1% over full range. |
| |
|
OPTICAL |
Target Range: |
14 in., (355.6 mm) to infinity. |
| Field of View: |
.382 degrees = .080 in. per ft. or
6.667 mm per m. |
| Target Size: |
TARGET (in.) = (separation + 4) / 150.
TARGET (mm) = (separation + 101.6) / 150. |
| |
|
Analog
|
|
Method: |
Continuously proportional, Industry
Standard current loop generator with internal loop power supply,
lower and upper range output limiters, and loop failure alarm. |
|
|
Range: |
0 - 20 mA, or 4 - 20 mA, operator
selectable. |
|
OUTPUTS |
Accuracy: |
+/- .15 %. |
|
|
Resolution: |
16 BIT, (.0015 %.). |
|
|
Termination: |
600 Ohms maximum loop resistance. |
|
Relay |
|
Self-Check Relay: |
1 Form C contact, 2 Amp, 30 VDC. |
|
|
Hi-Temp Alarm Relay |
2 Form C contacts, 2 Amp, 30 VDC. |
| |
|
OPERATING TEMPERATURE |
950S Viewing Head |
-40° to 55° C, -40° to 130° F No internal heater.
-55° to 55° C, -67° to 130° F Internal Heater, Option 1.
Upper limit increased to 85° C, 185° F with water jacket. |
| P905 Signal Processor |
-40° to 55° C, -40° to 130° F No heating or cooling
required. |
| |
|
CABLE |
P905 to 950S |
3 conductor, 16 gauge, with overall shield. 300 V,
PVC. |
| Optional Heater Cable |
3 conductor, 16 gauge, 300 V, PVC. Such as
Belden-M-8618 |